[ Care and Use ManUal ]BioSuite Columns 1I. IntroductIonThank you for choosing a Waters BioSuite™ column. Every column is indi-vidually tested to assu
[ Care and Use ManUal ]BioSuite Columns 2a. Mobile PhasePrevent air bubbles from entering the column during its installation, use, and storage since t
[ Care and Use ManUal ]BioSuite Columns 3the flow in this reversed flow direction until a few drops of mobile phase exit from the column. Turn off the
[ Care and Use ManUal ]BioSuite Columns 4IV. Waters BIosuIte Phenyl hIc columnsThis section contains recommended operating conditions and specifica-ti
[ Care and Use ManUal ]BioSuite Columns 5Note: When using a high viscosity buffer and ambient temperature, the maxi-mum flow rate may have to be reduc
[ Care and Use ManUal ]BioSuite Columns 6concentrations. Prior to chromatography, test to ensure the sample does not precipitate at the organic concen
[ Care and Use ManUal ]BioSuite Columns 76. Recommended Temperature: 4 - 60 °C. Reduce flow rate when operating at low temperatures (e.g., 4 °C) to a