[ Care and Use ManUal ]
BioSuite Columns 1
I. IntroductIon
Thank you for choosing a Waters BioSuite
column. Every column is indi-
vidually tested to assure maximum product quality. Waters BioSuite high
performance chemistries are dedicated to the isolation, analysis, and charac-
terization of proteins and peptides. For the latest list on ion-exchange, size
exclusion, hydrophobic interaction and reversed-phase go to www.waters.
com/lifesciences
contents
I. IntroductIon
a. Mobile Phase
b. Flow Direction
II. connectIng your column to the hPlc system
III. Waters BIosuIte Pc
18
and PPhenyl rPc columns
IV. Waters BIosuIte Phenyl hIc columns
V. Waters BIosuIte ultra hIgh resolutIon (uhr), hIgh
resolutIon (hr), and sIze exclusIon (sec) columns
VI. Waters BIosuIte Q-PeeK and sP-PeeK columns
VII. Waters BIosuIte deae and sP nP columns
VIII. Waters BIosuIte Q, deae, sP, and cm columns
BIosuIte columns
Przeglądanie stron 0
1 2 3 4 5 6 7 8

Podsumowanie treści

Strona 1 - BIosuIte columns

[ Care and Use ManUal ]BioSuite Columns 1I. IntroductIonThank you for choosing a Waters BioSuite™ column. Every column is indi-vidually tested to assu

Strona 2

[ Care and Use ManUal ]BioSuite Columns 2a. Mobile PhasePrevent air bubbles from entering the column during its installation, use, and storage since t

Strona 3

[ Care and Use ManUal ]BioSuite Columns 3the flow in this reversed flow direction until a few drops of mobile phase exit from the column. Turn off the

Strona 4

[ Care and Use ManUal ]BioSuite Columns 4IV. Waters BIosuIte Phenyl hIc columnsThis section contains recommended operating conditions and specifica-ti

Strona 5

[ Care and Use ManUal ]BioSuite Columns 5Note: When using a high viscosity buffer and ambient temperature, the maxi-mum flow rate may have to be reduc

Strona 6

[ Care and Use ManUal ]BioSuite Columns 6concentrations. Prior to chromatography, test to ensure the sample does not precipitate at the organic concen

Strona 7

[ Care and Use ManUal ]BioSuite Columns 76. Recommended Temperature: 4 - 60 °C. Reduce flow rate when operating at low temperatures (e.g., 4 °C) to a

Strona 8

[ Care and Use ManUal ]Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.comBioSuite Columns